SIGLENT Digital Multimeter SC1016 User’s Guide
WWW.SIGLENT.COM
SIGLENT Digital Multimeter SC1016 User’s Guide
2.Application of 16 Channels
(1) 2-Wired Application
[1]
(DCV/ DCI/ ACV/ ACI 2WR/ 4WR/ CAP/ FREQ/ DIODE/ CONT / TEMP)
Remarks
:
[1] CH1 to CH12 can be used to measure DCV /ACV/ 2WR/ 4WR/ CAP/ FREQ/ DIODE/ CONT / TEMP. CH13 to CH16 can only be used for current
measurements, less-than 2.4A
(2) 4-Wire Resistance Applications
[1]
3.Front Panel operations
Press and to enter the operating menu of Utility function, as the following diagram shows.
Table 1. Scanner Function Menu Description
Function Menu
Settings
Description
Mode
Scan/Step
Set the operation mode
Time
0ms~999.999s
Sets the duration between each scan loop (Scan mode) or between each scanned channel
(Step mode)
Cycles
Auto/Man
Sets the number of scan operations
Channel Setup
Sets the scanned channel range, measurement function, and measurement parameters
Start
On/Off
Start or stop scan operation
Exit
Exit the scanner function
1.Operation mode setup
•Scan: Measures all specified channel ranges (Channel MIN~MAX) for each trigger event. Time settings are applied between each scan for the whole channel range.
•Step: Measures a single channel in the specified range (Channel MIN~MAX) at each trigger event. Time settings are applied for each channel.
2.Time setup
Use the direction keys to set the duration between each scan loop (Scan mode) or between each scanned channel (Step mode)
3.Cycles
•Auto:
The instrument will scan specified channel circularly after the scan operation start and you should stop the operation manually.
•Manual: Sets the number of scan operations by direction keys. The range of the setting is from 1 to 999. After starting the scan operation the instrument will not
stop scanning until reaches cycle number.
Remarks: [1] To minimize voltage errors, the remote sense connections (CH7,CH8 etc..) should be made as close to the device-under-test (DUT) as possible.