SDG2000X_ServiceManual_SM0202X-E01B.doc
13
Chapter 1 Specifications
These specifications apply to the SDG2000X series Arbitrary Waveform
Generators. To verify that a generator meets the specifications, it must first
meet the following conditions:
⚫
The generator must have been operating continuously for thirty minutes
within the specified operating 18
℃
~ 28
℃
temperature.
⚫
The generator must be within the factory calibration interval of one year.
Frequency Characteristics
Parameter
Min.
Typ.
Max.
Unit
Condition
Resolution
1μ
Hz
Initial
accuracy
-1
+1
ppm
25
℃
-2
+2
ppm
0~40
℃
1
st
-year aging -1
+1
ppm
25
℃
10-year aging -3.5
+3.5
ppm
25
℃
Sine Characteristics
Parameter
Min.
Typ.
Max.
Unit
Condition
Frequency
1μ
120M
Hz
Harmonic
distortion
-65
dBc
0 dBm, 0~10 MHz (Included)
-60
dBc
0 dBm, 10~20 MHz (Included)
-55
dBc
0 dBm, 20~40 MHz (Included)
-50
dBc
0 dBm, 40~60 MHz (Included)
-45
dBc
0 dBm, 60~80 MHz (Included)
-40
dBc
0 dBm, 80~100 MHz (Included)
-38
dBc
0 dBm, 100~120 MHz (Included)
Total Harmonic
Distortion
0.075
%
0 dBm, 10 Hz ~ 20 kHz
Non-harmonic
spurious
-70
dBc
≤50 MHz
-65
dBc
>50 MHz
Square Characteristics
Parameter
Min.
Typ.
Max.
Unit
Condition
Frequency
1μ
25M
Hz
Rise/fall times
9
ns
10% ~ 90%, 1 Vpp, 5
0 Ω Load
Overshoot
3
%
100 kHz, 1 Vpp, 5
0 Ω Load
Summary of Contents for SDG2000X Series
Page 27: ...SDG2000X_ServiceManual_SM0202X E01B doc 27 ...
Page 31: ...SDG2000X_ServiceManual_SM0202X E01B doc 31 ...
Page 33: ...SDG2000X_ServiceManual_SM0202X E01B doc 33 ...
Page 36: ...SDG2000X_ServiceManual_SM 0202X E01B doc 36 ...
Page 40: ...SDG2000X_ServiceManual_SM 0202X E01B doc 40 ...
Page 57: ...SDG2000X_ServiceManual_SM0202X E01B doc 57 Oscilloscope 300 MHz Bandwidth SDS2304X ...