SDG2000X_ServiceManual_SM0202X-E01B.doc
47
Frequency Response Verification
This test checks if the amplitude flatness is within the spec range using a
Power Meter.
1. Turn on the generator and choose CH1 as the operating channel. Set the
Load to 5
0 Ω.
2. Connect the Power Meter, PC and generator as shown below.
3. Select Sine waveform of the generator and set the amplitude to 1 V and 2.5
V, frequency to the values listed below in sequence.
Output
Voltage
1 V
2.5 V
Frequency CH1 CH2
Spec Range
(dBm)
CH1 CH2
Spec Range
(dBm)
10 KHz
3.6794 ~ 4.2794
11.64 ~12.24
100 KHz
3.6794 ~ 4.2794
11.64 ~12.24
1 MHz
3.6794 ~ 4.2794
11.64 ~12.24
5 MHz
3.6794 ~ 4.2794
11.64 ~12.24
10 MHz
3.6794 ~ 4.2794
11.64 ~12.24
20 MHz
3.6794 ~ 4.2794
11.64 ~12.24
30 MHz
3.6794 ~ 4.2794
11.64 ~12.24
50 MHz
3.6794 ~ 4.2794
11.64 ~12.24
80 MHz
3.6794 ~ 4.2794
11.64 ~12.24
100 MHz
3.6794 ~ 4.2794
11.64 ~12.24
110 MHz
3.5794 ~ 4.3794
11.54 ~12.34
120 MHz
3.5794 ~ 4.3794
11.54 ~12.34
4. Remove the BNC cable to CH2 output and perform the same verification as