SIGLENT
6 SDG2000X User Manual
3. Basic Waveform Parameters Area
Shows the current waveform‘s parameters of each channel. Press Parameter
and select the corresponding softkey to highlight the parameter to configure.
Then use number keys or knob to change the parameter value.
4. Channel Parameters Area
Displays the load and output settings of the currently selected channel.
Load
----Value of the output load, as selected by the user.
Press Utility
→ Output → Load, then use the softkeys, number keys or knob to
change the parameter value; or continue pressing the corresponding output
key for two second to switch between High Impedance and 50Ω.
High Impedance: display HiZ.
Load: display impedance value (the default is 50Ω and the range is 50Ω to
100kΩ).
Note: This setting does not actually change the instrument‘s output impedance
of 50Ω but rather is used to maintain amplitude accuracy into different load
values.
Output
----Channel output state.
After pressing corresponding channel output control port, the current channel
can be turned on/off.
5. LAN Status Icon
SDG2000X will show different prompt messages based on the current network
status.
This mark indicates LAN connection is successful.
This mark indicates there is no LAN connection or LAN connection is
unsuccessful.
6. Mode Icon
SDG2000X will show different prompt messages based on the current mode.
This mark indicates current mode is Phase-locked.
Summary of Contents for SDG2042X
Page 10: ......
Page 45: ...SIGLENT SDG2000X User Manual 35 Figure 2 16 Setting the Bandwidth...
Page 123: ...SIGLENT SDG2000X User Manual 113 Figure 2 71 Phase Deviation Interface...
Page 138: ...SIGLENT 128 SDG2000X User Manual Figure 3 1 Generate a Sine Waveform...
Page 140: ...SIGLENT 130 SDG2000X User Manual Figure 3 2 Generate a Square Waveform...
Page 152: ...SIGLENT 142 SDG2000X User Manual Figure 3 9 Generate an AM Modulation Waveform...
Page 154: ...SIGLENT 144 SDG2000X User Manual Figure 3 10 Generate a FM Modulation Waveform...
Page 156: ...SIGLENT 146 SDG2000X User Manual Figure 3 11 Generate a PM Modulation Waveform...
Page 158: ...SIGLENT 148 SDG2000X User Manual Figure 3 12 Generate a FSK Modulation Waveform...
Page 160: ...SIGLENT 150 SDG2000X User Manual Figure 3 13 Generate an ASK Modulation Waveform...
Page 162: ...SIGLENT 152 SDG2000X User Manual Figure 3 14 Generate a PSK Modulation Waveform...
Page 164: ...SIGLENT 154 SDG2000X User Manual Figure 3 15 Generate a PWM Modulation Waveform...
Page 166: ...SIGLENT 156 SDG2000X User Manual Figure 3 16 Generate a DSB AM Modulation Waveform...