Programming Guide
16
RELATED COMMANDS
*CAL
Note:
Parameter/command SDG800 SDG1000 SDG2000X
SDG5000 SDG1000X
TST
no
yes
yes
yes
yes
3.1.10 WAI
DESCRIPTION
The *WAI (WAIT to continue) command, requires by the IEEE
488.2 standard, has no effect on the instrument, as the
signal generator only starts processing a command when the
previous command has been entirely executed.
COMMAND SYNTAX
*WAI
RELATED COMMANDS
*OPC
3.1.11 DDR
DESCRIPTION
The DDR? query reads and clears the contents of the device
dependent or device specific error register (DDR). In case of a
hardware failure, the DDR register specifies the origin of the
failure.
QUERY SYNTAX
DDR?
RESPONSE FORMAT
DDR <value>
<value> : = 0 to 65535
EXAMPLE
DDR?
Return:
DDR 0
The following table gives details
:
Bit
Bit Value
Description
15...14
Reserved
13
8192
Time-base hardware failure detected
12
4096
Trigger hardware failure detected
11
Reserved
10
Reserved
9
512
Channel 2 hardware failure detected
8
256
Channel 1 hardware failure detected