Programming Guide
35
3.13 Configuration Command
DESCRIPTION
Sets or gets the power-on system setting..
COMMAND SYNTAX
SCFG(Sys_CFG)<parameter>
<parameter>:= {DEFAULT, LAST}
QUERY SYNTAX
SCFG (Sys_CFG)?
RESPONSE FORMAT
SCFG <parameter>
EXAMPLE
Set the power-on system setting to LAST.
SCFG LAST
3.14 Buzzer Command
DESCRIPTION
Turns on or off the buzzer.
COMMAND SYNTAX
BUZZ(BUZZer) <parameter>
<parameter>:= {ON, OFF}
QUERY SYNTAX
BUZZ (BUZZer)?
RESPONSE FORMAT
BUZZ <parameter>
EXAMPLE
Turn on the buzzer.
BUZZ ON
3.15 Screen Save Command
DESCRIPTION
Turns off or sets screen save time (default unit is minutes).
COMMAND SYNTAX
SCSV (SCreen_SaVe) <parameter>
<parameter>:= {OFF, 1, 5, 15, 30, 60, 120, 300 }
QUERY SYNTAX
SCSV (SCreen_SaVe)?
RESPONSE FORMAT
SCreen_SaVe <parameter>