SDG5000 User Manual 18
To Set the Output Load
For the [CH1] and [CH2] connector on the front panel, the generator has a
built-in 50Ω series impendence. If the actual load does not match the set one,
the displayed amplitude and offset will be incorrect. This function is used to
match the displayed voltage with the expected one.
Steps for setting the load:
Press CH1 →load→ HighZ, to enter the following interface.
Please note that the load parameter shown on the down bottom is the default
setting when the power is on or the pre-set load value. If the current value is
valid for the output, then current value will be used.
Figure 1- 19 Set the Output Load
Instruction
SDG5000 Series have a fixed 50Ω Series Impendence. No matter what value
the set parameter is, if the real load is different from the set one, the displayed
voltage will not equal the real voltage.