SDG6000X Service Manual
15
WWW.SIGLENT.COM
mM
Chapter 1 Specifications
All specifications apply to both channels. Unless otherwise stated, all
specifications are not guaranteed unless the following conditions are met:
The generator has been working continuously for at least 30 minutes in a
specified temperature range (18
℃
~ 28
℃
)
The generator is within the valid calibration period.
Frequency Characteristics
Parameter
Min.
Typ.
Max.
Unit
Condition&Note
Resolution
1μ
Hz
Initial
accuracy
-1
+1
ppm
25
℃
-2
+2
ppm
0~40
℃
1
st
-year aging -1
+1
ppm
25
℃
10-year aging -3.5
+3.5
ppm
25
℃
Sine Characteristics
Parameter
Min.
Typ.
Max.
Unit
Condition&Note
Frequency
1
μ
500M
Hz
SDG6052x
1
μ
350M
Hz
SDG6032x
1
μ
200M
Hz
SDG6022x
Harmonic
distortion
-65
dBc
0 dBm, 0~1 MHz (Included)
-60
dBc
0 dBm, 1~60 MHz (Included)
-50
dBc
0 dBm, 60~100 MHz (Included)
-40
dBc
0 dBm, 100~200 MHz (Included)
-30
dBc
0 dBm, 200~300 MHz (Included)
-28
dBc
0 dBm, above 300 MHz
Total Harmonic
Distortion
0.075
%
0 dBm, 10 Hz ~ 20 kHz
Non-harmonic
spurious
-60
dBc
0 dBm,
≤350 MHz
-55
dBc
0 dBm, >350 MHz
Output Range
Note
2 m
20
Vpp
≤
40 MHz, HiZ load
2 m
10
Vpp
40 MHz ~ 120 MHz ( included ),
HiZ load
2 m
5
Vpp
120 MHz ~ 160 MHz ( included ),
HiZ load
2 m
3
Vpp
160 MHz ~ 350 MHz ( included ),
HiZ load
2 m
1.28
Vpp
above 350 MHz, HiZ load
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...