SDG6000X Service Manual
19
WWW.SIGLENT.COM
mM
PRBS
Sequence
length
2m-1, m = 3, 4, ... , 32
Rise/fall times
1 n
1 u
s
SDG6052X, SDG6032X.
10% ~ 90%, 1 Vpp, 50
Ω load
2 n
1 u
s
SDG6022X. 10% ~ 90%,
1 Vpp, 50
Ω load
Output Range
Note
2 m
20
Vpp
≤ 40 Mbps, HiZ load ,
2 m
10
Vpp
40 ~ 240 Mbps ( included ),
HiZ load
2 m
5
Vpp
Above 240 Mbps , HiZ load
Note :
The specification will be divided by 2 while applied to a 50
Ω load.
Output
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Accuracy
±(1%+1mVpp)
10 kHz sine, 0 V offset
Amplitude
flatness
-0.3
+0.3
dB
50Ω load, 0.5 Vpp, compare
to 1 MHz Sine
Output
impedance
49.5
50
50.5
Ω
100 kHz sine
Output current -200
200
mA
Crosstalk
-60
dBc
CH1=CH2=0 dBm, Sine, 50 Ω load
Protection
Current
limiting,
Over voltage protection
Current-limit
threshold
±200
mA
Overvoltage
protection
threshold
±3.5
±4
±4.5
V
The amplitude of the generator <3.2
Vpp and the DC offset <|2 VDC|
±10.5
±11
±11.5
V
The amplitude of the generator
≥
3.2 Vpp or the DC offset
≥
|2 VDC|
Modulation
AM
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Carrier
Sine, Square, Ramp, Arb
Modulation
source
Internal/External
Modulation
wave
Sine, Square, Ramp, Noise, Arb
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...