SDG6000X Service Manual
20
WWW.SIGLENT.COM
mM
Modulation
Modulation
depth
0
120
%
Modulation
frequency
1 m
1 M
Hz
While modulation source
is “Internal”
FM
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Carrier
Sine, Square, Ramp, Arb
Modulation
source
Internal/External
Modulation
wave
Sine, Square, Ramp, Noise, Arb
Frequency
deviation
0
0.5*BW
BW is the max. frequency. Limited
by frequency setting
Modulation
frequency
1 m
1M
Hz
While modulation source
is “Internal”
PM
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Carrier
Sine, Square, Ramp, Arb
Modulation
source
Internal/External
Modulation
wave
Sine, Square, Ramp, Noise, Arb
Phase
deviation
0
360
°
Modulation
frequency
1 m
1 M
Hz
While modulation source
is “Internal”
ASK
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Carrier
Sine, Square, Ramp, Arb
Modulation
source
Internal/External
Modulation
wave
Square with 50% duty cycle
Keying
frequency
1 m
1 M
Hz
While modulation source
is “Internal”
FSK
Parameter
Min.
Typ.
Max.
Unit
Condition & Note
Carrier
Sine, Square, Ramp, Arb
Modulation
source
Internal/External
Modulation
wave
Square with 50% duty cycle
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...