SDG6000X Service Manual
49
WWW.SIGLENT.COM
mM
substitute with the instrument of equivalent accuracy.
Instrument
Requirements
Recommended Model
Digital Multimeter (DMM)
DC volts
accuracy: 100 ppm
resolution: 100
μV
SDM3055
Power Meter
Frequency:
10 KHz to 6 GHz
Agilent U2004
Frequency Counter
Accuracy: 0.5 ppm
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations:
Updating SDG’s firmware to the newest version is strongly recommended.
Assure that the calibration ambient temperature is stable and between
18
℃
and 28
℃
. Ideally, the calibration should be performed at 23
℃
±1
℃
.
Assure ambient relative humidity is less than 80%.
Allow a 30-minute warm-up period before verification or adjustment.
Keep the measurement cables as short as possible, consistent with the
impedance requirements.
Use a 5
0 Ω coaxial cable.
Performance Verification Test
The performance verification tests are recommended as acceptance tests
upon receipt of the instrument. The acceptance test results should be
compared to the specifications given in chapter 1. After acceptance, repeat the
performance verification tests within every calibration interval.
If the instrument fails performance verification, adjustment or repair is required.
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...