SDG6000X Service Manual
54
WWW.SIGLENT.COM
mM
General Adjustment Procedure
The following explains how to adjust the SDG6000X series generator for
optimum operating performance.
Channel Adjustment
which includes four steps: feedback channel
adjustment, channel self adjustment, vertical accuracy calibration and
frequency counter calibration.
Feedback Channel Adjustment
which acts as a standard calibration
module, providing the accuracy assurance for Self Adjust.
Channel Self Adjustment
which includes internal control channel
adjustment and signal channel adjustment.
Vertical Accuracy Calibration
which
calibrates the amplitude at 10 kHz
frequency.
Frequency Counter Calibration
which
calibrates the trigger level of the
counter.
Frequency Response Adjustment
which could properly compensate the
amplitude as the frequency increases.
TimeBase Calibration
which
calibrates the output frequency accuracy.
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...