SDG6000X Service Manual
57
WWW.SIGLENT.COM
mM
Feedback Channel Adjustment
The internal feedback channel receives the feedback signal from all the other
analog channels when perform self adjustment. This adjustment must be
performed prior to Self Adjust to make sure it is performed based on the best
adjusting accuracy.
1. Connect the DMM, PC and SDG6000X generator as shown below:
2. Double click and run the script
FeedbackAdjust.py
prepared previously.
The CH1 will be automatically turned on upon running of the script.