Quick Guide for General Failures
The general hardware failures are described in the following. This information
may help in determining the cause of some general hardware failures:
1. No start-up after pressing the power button:
(1) Check if the power cord is correctly connected.
(2) Check if the power button is operating correctly.
(3) Check whether the fuse has burned out. If the fuse is blown, please
replace with a fuse of the same rating.
(4) Check if the connection between the power supply and the channel board
is tight.
(5) If the instrument still does not work normally, please contact SIGLENT.
2. Starts up with a dark screen:
(1) Check the connection between the power supply and the channel board.
(2) Check the connection between the keypad circuit board and the channel
board.
(3) If the instrument still does not work normally, please contact SIGLENT.
3. No response after pressing any button, or abnormal display of the
screen:
(1) Check the connection between the keypad circuit board and the channel
board.
(2) If the instrument still does not work normally, please contact SIGLENT.
4. The output voltage amplitude measured is higher or lower than
expected:
(1) Check the connection between the generator and the load.
(2) Check if the impedance set in generator matches the input impedance of
the load.
(3) If the instrument still does not work normally, please contact SIGLENT.
Summary of Contents for SDG6000X Series: SDG6022X
Page 1: ...SDG6000X Series Pulse Arbitrary Waveform Generator Service Manual SM0206X E01A 06X E01A...
Page 2: ......
Page 33: ...SDG6000X Service Manual 31 WWW SIGLENT COM mM...
Page 37: ...SDG6000X Service Manual 35 WWW SIGLENT COM mM...
Page 39: ...SDG6000X Service Manual 37 WWW SIGLENT COM mM...
Page 42: ...SDG6000X Service Manual 40 WWW SIGLENT COM mM...
Page 46: ...SDG6000X Service Manual 44 WWW SIGLENT COM mM...
Page 75: ...SDG6000X Service Manual 73 WWW SIGLENT COM mM...