SDM3055/SDM3055A 38
2.6.3 CONFigure:CURRent:{AC|DC} [{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for AC or
DC current measurements. Also specifies the range and resolution through the incoming
parameters.
Parameter
Typical Return
<range>
:
{200uA|2mA|20mA|200mA|2A|10A|AUTO}
,
default AUTO (autorange),The 200uA and 2mA file can be
set only in DC current mode , The minimum gear for 20mA
in AC mode .
(none)
Using 2 a range configuration ac current measurements. Two measurement and
reading:
CONF:CURR:AC 2
SAMP:COUN 2
READ?
Typical Response
:
+4.32133675E-04,+4.18424606E-04
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
Use READ? or INITiate to start the measurement.