SDM3055/SDM3055A 48
2.8.2 MEASure:CURRent:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for AC or
DC current measurements and immediately triggers a measurement. Also specifies the
stalls through the incoming parameters.
Parameter
Typical Return
<range>
:
{200uA|2mA|20mA|200mA|2A|10A|AUTO}
,
default AUTO(autorange).The 200uA and 2mA file can only
be set in DC current mode,AC mode minimum gear for
20mA
None
Configure AC current measurement using the 2A range.Make and read two
measurements:
CONF:CURR:AC 2
SAMP:COUN 2
READ?
Typical Response
:
+4.32133675E-04,+4.18424606E-04
You can let autoranging select the measurement range, or you can manually select
a fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a <
resolution
>
because the instrument cannot accurately resolve the integration time (especially if
the input continuously changes). If your application requires autoranging, specify
DEFault for the <
resolution
> or omit the <
resolution
> altogether.
If the input signal is greater than can be measured on the specified manual range,
the instrument displays the word Overload on front panel and returns "9.9E37" from
the remote interface.
Summary of Contents for SDM3055A
Page 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Page 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 113: ......
Page 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...