SDM3055/SDM3055A 56
2.9.2 [SENSe:]CURRent:{AC|DC}:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]CURRent:{AC|DC}:NULL:VALue? [{MIN|MAX|DEF}]
Sets the null value for AC or DC current measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
-12 to 12 A, default 0
+1.00000000E-01
Configure AC current measurements, using the null function to subtract 100 mA
from the measurements. Then
make two measurements and send them to the instrument's output buffer:
CONF:CURR:AC
CURR:AC:NULL:STAT ON;VAL 100mA
SAMP:COUN 2
READ?
Typical Response
:
-9.92150377E-02,-9.98499843E-02
Specifying a null value disables automatic null value selection
([SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO OFF).
To use the null value, the null state must be on
([SENSe:]CURRent:{AC|DC}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.