[SENSe:]{RESistance|FRESistance} Subsystem
This subsystem configures two- and four-wire resistance measurements.
Command Summary
[SENSe:]{RESistance|FRESistance}:NPLC
[SENSe:]{RESistance|FRESistance}:NULL[:STATe]
[SENSe:]{RESistance|FRESistance}:NULL:VALue
[SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO
[SENSe:]{RESistance|FRESistance}:RANGe
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO
2.9.13 [SENSe:]{RESistance|FRESistance}:NPLC {<PLC>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for all alternating current
measurements.Integration time is the period that the instrument's analog-to-digital (A/D)
converter samples the input signal for a measurement.
Parameter
Typical Return
{0.3|1|10}
,
default 1
On the front panel, 0.3|1|10 corresponds to the
Speed menu under fast|middle|slow
+1.0001
Configure 4-wire resistance measurements using a 10 PLC integration time.
CONF:RES
CURR:DC:NPLC 10
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).