2.9.15 [SENSe:]{RESistance|FRESistance}:NULL:VALue
{<value>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for all resistance measurements.
Parameter
Typical Return
-120 MΩ to +120 MΩ
,
default
为
0
+1.0002
Configure 2-wire resistance measurements, using the null function to remove
100 mΩ of wiring resistance. Make and read two measurements:
CONF:RES
RES:NULL:STAT ON;VAL 0.1
SAMP:COUN 2
READ?
Typical Return
:
+1.0402,+1.0402
Specifying a null value disables automatic null value selection
([SENSe:]{RESistance/FRESistance}:NULL:VALue:AUTO OFF).
To use the null value, the null state must be on
([SENSe:]{RESistance|FRESistance}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.