2.9.16 [SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO
{ON|1|OFF|0}
[SENSe:]{RESistance|FRESistance}:NULL:VALue:AUTO?
Enables or disables automatic null value selection for all resistance measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default on
0 (OFF) or 1 (ON)
Using automatic selection of zero value to measuret
:
RES:NULL:VAL:AUTO ON
READ?
Typical Return
:
+1.2302,+1.1602
When automatic reference selection is ON, the first measurement made is used as
the null value for all subsequent measurements.
[SENSe:]{RESistance/FRESistance}:NULL:VALue will be set to this value. Automatic
null value selection will be disabled.
When automatic null value selection is disabled (OFF), the null value is specified by
this command: [SENSe:] {RESistance/FRESistance}:NULL:VALue.
The instrument enables automatic null value selection when the null function is
enabled ([SENSe:] {RESistance/FRESistance}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.