SIGLENT Remote Manual
75 SDM3055/SDM3055A
2.9.22 [SENSe:]TEMPerature:TRANsducer?
Query user-defined/manufacturer curing RTD/THER sensor list.
Parameter
Typical Return
(none )
MANU DEFINE,THER,BITS90
Query current effective sensor:
TEMP:TRAN?
Typical Response
:
PT100
2.9.23 [SENSe:]TEMPerature:{UDEFine|MDEFine}
:{THER|RTD}:TRANsducer <transducer>
Set the current effect of sensors.
Parameter
Typical Return
PT100(RTD)/{BITS90|EITS90|JITS90|KITS90|NIT
S90|RITS90|SITS90|TITS90}(THER)
(none)
Set the THER KITS90 as the current sensor
:
TEMP:MDEF:THER:TRAN KITS90
The information can be obtained through the
[SENSe:]TEMPerature:{UDEFine|MDEFine}:{THER|RTD}:TRANsducer:LIST?
This command will respond to the defaultsensor(KITS90) after a Factory Reset
(*RST), Instrument Preset (SYSTem:PRESet), or CONFigure function.