SDM3055/SDM3055A 80
2.9.28 [SENSe:]VOLTage:{AC|DC}:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]VOLTage:{AC|DC}:RANGe? [{MIN|MAX|DEF}]
Manually choose measurement range
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
<range>
:
{200 mV|2 V|20 V|200 V|1000
V(DC)/750V(AC)}
AC default:20V.
DC default:1000V.
+2.0000
Configure AC voltage measurements using the 100 V range. Make and read two
measurements:
CONF:VOLT:AC
VOLT:AC:RANG 2
SAMP:COUN 2
READ?
Typical Response
:
+8.21650028E-03,+8.17775726E-03
Selecting a fixed range ([SENSe:]<
function
>:RANGe) disables autoranging.
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
The instrument is set to the default range, with autoranging enabled
([SENSe:]VOLTage:{AC|DC}:RANGe:AUTO ON), after a Factory Reset (*RST) or an
Instrument Preset (SYSTem:PRESet)