SDM3055/SDM3055A 92
2.11 TRIGger Subsystem
The TRIGger subsystem configures the triggering that controls measurement acquisition.
Command Summary
TRIGger:COUNt
TRIGger:DELay
TRIGger:DELay:AUTO
TRIGger:SLOPe
TRIGger:SOURce
2.11.1 TRIGger:COUNt {<count>|MIN|MAX|DEF|INFinity} TRIGger:COUNt?
[{MIN|MAX|DEF}]
Selects the number of triggers that will be accepted by the instrument before returning to
the "idle" trigger state.
Parameter
Typical Return
1 to 1,000,000 or continuous (INFinity).
Default 1
+1.0000
For a continuous trigger (INFinity),
the query returns
"9.9E37".
Return ten sets of five DC voltage measurements,
CONF:VOLT:DC
SAMP:COUN 5
TRIG:COUN 10
READ?
Typical Response: -1.85425399E-04, … (50 measurements)
You can use the specified trigger count in conjunction with a sample count
(SAMPle:COUNt), which sets the number of samples to be made per trigger. In this
case, the number of measurements returned will be the sample count multiplied by
the trigger count.
You can store up to 1,000 measurements in the reading memory of the SDM3055x. If
reading memory overflows, new measurements will overwrite the oldest
measurements stored; the most recent measurements are always preserved. No
error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Questionable
Data Register's condition register (see Status System Introduction).
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).
Summary of Contents for SDM3055A
Page 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Page 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Page 113: ......
Page 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...