SDS1000X User Manual
115
To Make Automatic Measurement
Perform the steps below and select voltage or time parameters to make automatic
measurement.
1. Press the
Measure
button on the front panel to enter the MEASURE function menu.
At the same, the frequency and period are enabled with the current trigger channel,
the statistics also was enabled.
2. Press the
Source
softkey, and then use the
Universal Knob
to select the desired
channel. Only analog channels (CH1 and CH2) that are displayed are available for
measurements.
3. To select and display measurement parameters. Press the
Type
softkey, and then
turn the
Universal Knob
to select the desired measurement parameter.
4. Press the
Universal Knob
to add the measurement parameter, the parameters and
value will be shown above the menu, and the statistics status will update.
5. To turn off the statistic function, press the
Statistics
softkey to select
Off
.
The measurement display area can display 5 measurement parameters at most, and the
measurements will arrange according to the selecting order. If add a sixth measurement
parameter, it will delete the first measurement.
Figure 64 Select the Measurement Parameter
www.biall.com.pl
Summary of Contents for SDS1102X
Page 2: ...www biall com pl...
Page 19: ...SIGLENT SDS1000X User Manual XVII www biall com pl...
Page 20: ...www biall com pl...
Page 89: ...SIGLENT SDS1000X User Manual 69 Figure 28 Relative Window Trigger www biall com pl...
Page 91: ...SIGLENT SDS1000X User Manual 71 Figure 29 Interval Trigger www biall com pl...
Page 95: ...SIGLENT SDS1000X User Manual 75 Figure 32 Runt Trigger www biall com pl...
Page 109: ...SIGLENT SDS1000X User Manual 89 Figure 44 SPI Trigger www biall com pl...