SDS1000X User Manual
To Update Firmware and Configuration
The firmware and configuration can be updated directly via USB flash driver.
Do the following steps to update the firmware:
1. Insert the USB flash driver which with the firmware and the configure files inside into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Firmware
softkey and it will pop-out the message ―
Press „Single‟ to
continue and press „Run/Stop‟ to exit!
‖
7. Press the
Single
button to enter the SAVE/RECALL file system.
8. Turn the
Universal Knob
to select the update file which should be with an ADS postfix;
and then press the
Load
softkey to start update the firmware. The process needs
about 7 minutes. And during the update, do not cut off the oscilloscope, otherwise the
oscilloscope will not restart again.
9. After finish the update, the screen will pop-out the message ―
Update success, please
restart the DSO
‖, and you will heard a sound of buzzer.
10. Restart the oscilloscope to finish the firmware update.
Do the following steps to update the configuration:
1. Insert the USB flash driver which with the firmware and the configure files inside into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Configure
softkey to enter the SAVE/RECALL file system.
7. Turn the
Universal Knob
to select the update file which should be with an CFG
postfix; and then press the
Load
softkey to start update the firmware. The process
needs about 30 seconds.
8. After finish the update, the screen will pop-out the message ―
Update success, please
restart the DSO
‖, and you will heard a sound of buzzer.
9. Restart the oscilloscope to finish the configuration update.
www.biall.com.pl
Summary of Contents for SDS1102X
Page 2: ...www biall com pl...
Page 19: ...SIGLENT SDS1000X User Manual XVII www biall com pl...
Page 20: ...www biall com pl...
Page 89: ...SIGLENT SDS1000X User Manual 69 Figure 28 Relative Window Trigger www biall com pl...
Page 91: ...SIGLENT SDS1000X User Manual 71 Figure 29 Interval Trigger www biall com pl...
Page 95: ...SIGLENT SDS1000X User Manual 75 Figure 32 Runt Trigger www biall com pl...
Page 109: ...SIGLENT SDS1000X User Manual 89 Figure 44 SPI Trigger www biall com pl...