The display of waveform is ladder-like:
1) The horizontal time base might be too low. Increase the horizontal time base to
increase the horizontal resolution and improve the display.
2) If the display
Type
is ―Vectors‖, the lines between the sample points may cause
ladder-like display. Set
Type
to ―Dots‖ to solve the problem.
7.
Fail to connect PC through USB:
Check the
IO Setting
in
Utility
to make sure whether the setting in
USB Device
matches the device currently connected. If needed, restart the oscilloscope.
8.
The USB storage device cannot be recognized:
1) Check whether the USB storage device can work normally.
2) Make sure whether the USB interface can work normally.
3) Make sure that the USB storage device being used is flash storage type. This
oscilloscope does not support hardware storage type.
4) Make sure whether the system of the USB storage device is FAT32.
5) Restart the instrument and then insert the USB storage device to check it.
6) If the USB storage device still cannot be used normally, please contact
SIGLENT
.