SDS1000X User Manual
Sample Rate
The maximum sample rate of SDS1000X is 1GSa/s. The actual sample rate of the
oscilloscope is determined by the horizontal scale. Turn the
Horizontal Scale Knob
to
adjust the sample rate.
The actual sample rate is displayed in the information area at the upper- right corner of the
screen.
The influence on the waveform when the sample rate is too low:
1. Waveform Distortion:
when the sample rate is too low, some waveform details are
lost and the waveform displayed is rather different from the actual signal.
2. Waveform Confusion:
when the sample rate is lower than twice the actual signal
frequency (Nyquist Frequency), the frequency of the waveform rebuilt from the sample
data is lower than the actual signal frequency. The most common aliasing is the jitter
on fast edge.
3. Waveform Leakage:
when the sample rate is too low, the waveform rebuilt from the
sample data does not reflect all the actual signal information.
Pulse disappeared