SDS1000X User Manual
43
Peak Detect
In this mode, the oscilloscope acquires the maximum and minimum values of the signal
within the sample interval to get the envelope of the signal or the narrow pulse of the
signal that might be lost. In this mode, signal confusion can be prevented but the noise
displayed would be larger.
In this mode, the oscilloscope can display all the pulses with pulse widths at least as wide
as the sample period.
Figure 15 Pulse With 0.1% Duty, Normal Mode
Figure 16 Pulse With 0.1% Duty, Peak Detect Mode