Note: Press the
Auto Setup
button will set the trigger type to Edge and slope to rising.
Edge Trigger
Edge trigger distinguishes the trigger points by seeking the specified edge (rising, falling,
rising & falling) and trigger level.
Rising Edge Falling Edge
Trigger Level
Trigger Point Trigger Point
1. Press the
Setup
button on the front panel to enter the TRIGGER system function
menu.
2. Press the
Type
softkey; turn the
Universal Knob
to set select
Edge
and then push
the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select
CH1
,
CH2
,
EXT
,
EXT/5
or
AC Line
as the trigger source.
4. Press the
Slope
softkey; turn the
Universal Knob
to select the desired trigger edge
(rising, falling or rising & falling), and then press down the knob to confirm. The current
trigger slope is displayed at the upper right corner of the screen.
5. Turn the
Trigger Level Knob
to adjust the trigger level to obtain stable trigger.
Figure 23 Edge Trigger