SDS1000X User Manual
Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another as shown in the
picture below.
Positive runt pulse
High level
Low level
Negative runt pulse
A positive runt pulse across through a lower threshold but not an upper threshold.
A negative runt pulse across through an upper threshold but not a lower threshold.
To trigger on runt pulse:
1. Press the
Setup
button on the front panel to enter the
TRIGGER
system function
menu.
2. Press the
Type
softkey; then turn the
Universal Knob
to select
DropOut
and push
down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the
trigger source.
4. Press the
Polarity
softkey to select
Positive
or
Negative
pulse to trigger.
5. Press the
Limit Range
softkey; turn the
Universal Knob
to select the desired
condition (
<
,
>
,
<>
or
><
).
6. Press the
Time Setting
softkey, and then turn the
Universal Knob
to select the
desired value.
7. Press the
Next Page
softkey to enter the second page of the TRIGGER system
function menu. Press the
Lower Upper
softkey to select
Lower
or
Upper
trigger level,
and the turn the
Universal Knob
to set the position.
www.biall.com.pl
Summary of Contents for SDS1102X
Page 2: ...www biall com pl...
Page 19: ...SIGLENT SDS1000X User Manual XVII www biall com pl...
Page 20: ...www biall com pl...
Page 89: ...SIGLENT SDS1000X User Manual 69 Figure 28 Relative Window Trigger www biall com pl...
Page 91: ...SIGLENT SDS1000X User Manual 71 Figure 29 Interval Trigger www biall com pl...
Page 95: ...SIGLENT SDS1000X User Manual 75 Figure 32 Runt Trigger www biall com pl...
Page 109: ...SIGLENT SDS1000X User Manual 89 Figure 44 SPI Trigger www biall com pl...