SPS5000X series service manual
Figure 3-2
Calibration instrument: high precision multimeter, electronic load.
SCPI calibration command description
1)
CALibrate:STATe <CHn>,<State>
This command is used to set the calibration mode status.
CHn:
CH1 Select to calibrate channel 1
CH2 Select to calibrate channel 2
CH3 Select to calibrate channel 3
State:
1 Calibration mode is on
0 Calibration mode is off
E.g: CALibrate:STATe CH1,1
2)
CALibrate:CLEAn <CHn>,<type>
This command is used to clear the specified type of calibration data.
type:
VSET Setting voltage
VMEAS Measuring voltage
CSET Setting current
CMEAS Measuring current
3)
CALibrate:MEAS:<type> <CHn>,<Start>,<End>,<X1>,<Y1>,<X2>,<Y2>
This command is used to calibrate readback voltage.