10 SSG5000X Service Manual
USB Device Test
To test if the USB Device interface is operating normally.
Tools
:
● A computer with USB interface that is compatible with running National
Instruments NI-MAX software
● A standard USB cable (Type A-B)
● NI-MAX software
Steps
:
1. Download and install National Instruments NI MAX software by following the
installation instructions provided by National Instruments.
2. Connect the signal generator USB Device port and the computer using an
USB cable.
3. Run NI MAX
software. Click “Device and interface” at the upper left corner of
the NI software interface
and immediately displays the “USBTMC” device
symbol.
4. Click “Open VISA Test Panel” option button, and then the following
Interface will appear. Then click the
“Input/Output” option button and click
the “Query” option button in order to view the Read operation information.