16 SSG5000X Service Manual
10 MHz output accuracy test
Specification
Reference frequency
10.000000 MHz
Initial calibration accuracy
<0.2 ppm
Test Connection Diagram
Test Procedures
1. Connect ref out port of the signal generator to the channel A of the
SDG6000X. The SDG6000X should be connected to an external OCXO 10
MHz reference oscillator.
2. Set the SDG6000X to frequency counter mode, and set frequency ref to
10.000000 MHz
3.
Check if the frequency deviation ≤ 0.2 ppm
Test Record Form
Frequency
Frequency Deviation
Pass/Fail
10.000000 MHz