SIGLENT
34 SSG5000X Service Manual
LF AC test
Specification
Internal modulation generator(LF)
Frequency response
Sine wave < 0.3 dB
Test Devices
1. Power Sensor × 1
2. N-BNC Cable × 1
3. USB Cable × 1
Test Connection Diagram
Test Procedures
1. Connect the LF output terminal of the SSG5000X with the input terminal of
the power sensor. Connect the
[USB
]
connector of the SSG5000X
with the USB interface of the power sensor as shown in figure above.
2. Enable LF output switch. Select Sine waveform of the generator and set the
amplitude to 0 dBm and then step through the frequencies listed below in
sequence. Then, set the amplitude to -20 dBm and repeat the frequency
steps.
Output
Voltage
0 dBm
-20 dBm
Frequency
Power Meter
Measurement
Value
Spec Range
(dBm)
Power Meter
Measurement
Value
Spec Range
(dBm)
100 kHz
-0.3 ~ 0.3
-20.3 ~ -19.7
200 kHz
300 kHz
400 kHz
500 kHz
600 kHz
700 kHz