SIGLENT
36 SSG5000X Service Manual
LF DC test
Specification
DC Offset error
±(1%+3 mV)
Test Devices
1. Digital Multimeter × 1
2. Banana-to-single-BNC Cable × 1
Test Connection Diagram
Test Procedures
1. Connect the LF output terminal of the SSG5000X with the Voltage input
terminal of the Digital Multimeter as shown in figure above.
2. Enable LF output switch. Select DC waveform of the generator and set the
instrument to each output value described in the table below and measure
the output voltage with the DMM.
DC Offset
DMM Measurement Value
Spec Range
± (1%+3 mV)
-2 V
-2.017 V ~ -1.983 V
-1 V
-1.007 V ~ -0.993 V
-100 mV
-102 mV ~ -98 mV
-10 mV
-12.9 mV ~ -7.1 mV
0 V
-3 mV ~ 3 mV
10 mV
6.9 mV ~ 13.1 mV
100 mV
96 mV ~ 104 mV
1 V
0.987 V ~ 1.013 V
2 V
1.977 V ~ 2.023 V
Note
: The output impedance of generator is 50Ω and the input impedance of DMM is high-z, so the
measured value should divide by 2 to get a accurate value.
3. Compare the measured voltage with the spec range shown in the table
above.