6.4 Resistance Test, 4-wire
The following procedure may be used to verify the accuracy of the 4-wire function.
1. If you have not done so, install the SMU2060/64 DMM and place the covers back on to the computer. Ensure
that the computer has been on for at least one-half hour, with the covers on, before conducting this test.
2. Connect the DMM
V,
+ & -
terminals to the calibrator HI & LO Output. Connect the DMM’s
I, 4W
+ & -
terminals to the HI & LO Sense terminals.
3. Select the 4W
function on the DMM, Autorange. Set the calibrator to 0
. Be certain that the calibrator is set
to external sense ("EX SNS" on the Fluke 5700A or “4-Wire Comp” on the 5520A). Allow the DMM to settle for
a few seconds, and perform the
Relative
function.
4. Apply the following Resistance values to the
V,
+ & -
terminals. Check to see that the displayed reading on
the DMM is within the indicated range.
Table 9-4 Resistance Test, 4-wire
Step
Range
Input
Minimum Reading
Maximum Reading
1
24.000000
00.000000
00.000000
00.001000
2
24.000000
10.000000
09.98200
10.001800
3
240.00000
000.00000
000.00000
000.00500
4
240.00000
100.00000
099.98800
100.01200
5
2.4000000 k
0.0000000 k
0.0000000
k
000.00003
k
6
2.4000000 k
1.0000000 k
0.9999070
k
1.0000930
k
7
24.000000 k
00.000000 k
00.000000
k
00.000350
k
8
24.000000 k
10.000000 k
09.999050
k
10.000950
k
9
240.00000 k
000.00000 k
000.00000
k
000.00500
k
10
240.00000 k
100.00000 k
099.98800
k
100.01200
k
11
2.4000000 M
0.0000000 M
0.0000000
M
0.0000700
M
12
2.4000000 M
1.0000000 M
0.9995300
M
1.0004700
M
13
24.0000 M
00.0000 M
00.0000
M
00.0006
M
14
24.0000 M
10.0000 M
00.0998
M
10.0206
M
Note 1: Some ranges apply to 2064 only. Please refer to chapter 2.0 (Specification).
Note 2: The use of 4-wire Ohms for resistance values above 300 k
is not recommended.