Document Reference: X-0036-01-07 Cadmus & Ratifi Operating Manual
Drawn: 13/11/2020
Page 24 of 66
1.13.4
Cadmus-Config basics
Opening Cadmus-Config will present you with the screen below. Only two tabs are available, Device and Release. Release shows information about the current
application version. Device is used to make connection to the device.
Select “Connect” to make the USB connection active. Note: connection may fail if Windows is currently installing the
default driver; in this case you just need to wait a
few moments until Windows has completed the task.
If your device display matches the image to the right. The real time clock is not
currently set. Cadmus-
Config will display the warning “Real Time Cloc
k is out
of sync, continue to set clock”. Select “OK” and the clock with automatically
set and the device re-
booted. You will be taken back to the “Connect” screen.
Please press “Connect” again.
This normally only occurs when a device is new and has been without power.
The device will keep the real time clock running for up to 30 minutes with no
power connected, for when battery replacements are required.