Instruction Manual DualScat Ex
10118E/5
27
Action
1.
Disconnect the power supply to the control
unit.
2.
Wait 10 minutes before continuing.
Allows instrument to cool
to a safe temperature level
and residual charges to
dissipate.
3.
Unscrew socket head screw A by about
½ turn and remove the locking ring.
4.
Unscrew nut B far enough to be able to
swing the locking strap easily. Now remove
the closure clamp and then the enclosure.
5.
Remove the desiccant and dispose of it as
directed in the disposal table in Section 9.
6.
Remove the new desiccant from it protec-
tive pouch and insert it as shown in the ad-
jacent illustration (Spares
Section 10).
Press the desiccant all the way into the
holder, so that it cannot come into contact
with the enclosure.
Reclose the enclosure immediately to
prevent the desiccant from absorbing ambi-
ent moisture.
The cable gland must match the flat
side of the mounting plate.
7.
Put on the closure clamp, swing the locking
strap back in place, and retighten the nut B.
Replace desiccant
A
B
B