Instruction Manual DualScat Ex
10
10118E/5
Important information for photometer installation:
Scratches on the blanking glass can lead to incorrect measurements:
Ensure
that no scratches occur on the blackened outer face and uncolored inner face
when installing and removing the blanking glass. Always handle the blanking
glass with care.
The outer face of the blanking glass is blackened with a PVD chrome coating.
The inner side of the blanking plate is sand blasted.
The photometer must be installed in the pipe at a point at least 2 m away from
inspection windows or other sources of stray light.
3.1.2
Installation of the SIREL SMD Control unit
The SIREL SMD and any peripheral devices may not be installed or operated in
explosive atmospheres.
The SIREL SMD can be mounted right on a wall, on a mounting grate, or on a
stand.
Depending on the cable used (cross-section), the maximum distance between
control unit and photometer is given in the following table:
Cable cross-section
[mm
2
]
Max. distance
SIREL SMD [m]
Max. distance
SITRA [m]
0.14
42
35
0.25
75
65
0.34
103
90
0.50
148
130
0.75
223
200
standard version
1.00
289
260
1.50
414
370
Refer to Section 0 for a detailed dimension drawing of the SIREL SMD.
Summary of Contents for DualScat Ex
Page 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Page 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Page 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Page 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...