Instruction Manual OilGuard 2 W
Mounting
24
14809E/2
The following procedure describes how to mount the sample connections on the
OilGuard 2 W:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the optics unit according to
Section 9.1.1 and fasten on the docking sta-
tion.
Warning! UV radiation.
Use UV goggles and gloves.
2.
Fasten the inlet hose on the inlet pipe
(Figure 3, pos. 1).
3.
Fasten the outlet hose on the outlet pipe
(Figure 3, pos. 7).
When doing so, press the outlet cone
(Figure 3, pos. 5) downward by hand so that
the outlet pipe (Figure 3, pos. 7) is not pressed
into the measuring cell space.
Flooding of the measuring cell
due to improper mounting of the
outlet hose.
The outlet hose must be fastened
so that the holes on the side are
not covered (flood protection).
Section 4.4.1
4.
Check to ensure the inlet pipe (Figure 3,
pos. 1) projects 5 mm out of the holder.
5.
Check to ensure the outlet cone (Figure 3, pos.
5) is in contact with the conduit gland
(Figure 3, pos. 6).
6.
Place the optics unit back on the measuring
cell unit and lock with the four mounting clips.
4.5
Mounting the optional accessories
4.5.1
General information on the optional accessories
CAUTION!
Note the following when mounting the optional accessories:
All hose connections must be secured with hose clamps. The connections should be
checked about two weeks after installation to ensure that no air can be pulled in.
A minimum flow of 1 l/min. must be guaranteed on the overflow of the level control.
The dimensions on the dimension sheets must be adhered to (course and radii of the
hoses, spacing etc.).
4.5.2
Mounting the level control
Mounting of the level control is made according to the dimension sheets
OILGUARD 2 W/1-MB
.