Operation
Instruction Manual OilGuard 2 W
14809E/2
45
7.10
Switching to service mode
The system is configured in service operation. The measuring procedure is interrupted and
the main menus appear on the display. Service operation is accessed as follows:
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
The main menus appear.
The instrument is now in service op-
eration.
The following applies in service operation:
* The measuring values remain on the last values on the digital interfaces.
* Depending on the configuration, the current outputs go to 0/4 mA or remain on the
last measuring values.
The limits are deactivated.
If an output for service is programmed, it is activated.
Fault messages are suppressed.
* This does not apply when the
Current outputs\General\For service
parameter is set to
Measure
.
For measuring operation, press the
Meas
button. When switching from service operation
to measuring operation, an hourglass appears in the information bar for about 10 seconds.
The measuring values are frozen during this time.