Settings
Instruction Manual OilGuard 2 W
14809E/2
51
8.3
Setting the limits
The limits have to be configured accordingly so that they are not only displayed, but that the
outputs are also switched. Section 8.4
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Limits
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
4.
Select between
L1 .. n
.
5.
Select the
Source
.
The following selection is available
(when present):
C1
(measuring channel)
A1 .. n
(analog channels)
6.
Define the
Mode
.
The following selection is available:
Inactive
(limit monitoring of
this channel is deactivated).
Exceeded
(limit active when
the set threshold value is ex-
ceeded).
Undershot.
(limit active when
the set threshold value is under-
shot).
7.
Define the upper limit, lower limit, cut-in delay
and cut-out delay with the number pad.
Pressing the current number
value takes you to the entry mode.
8.
Press the
Meas
button.
The instrument is in measuring op-
eration again.