Instruction Manual OilGuard 2 W
Servicing
60
14809E/2
9.1.2
Checking the flow rate and cleanliness
The following procedure describes how the flow rate is checked:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the optics unit on the photometer
according to Section 9.1.1 and fasten on the
docking station.
Danger due to UV radia-
tion.
Use UV goggles and gloves.
2.
Check the flow rate of the sample according
to the technical data. Section 2.4
3.
Check the water jet.
Normal jet, reflections are visible in the
water jet.
Uneven water
jet after water
stop
Too little water
Too much water
With large air
bubbles
4.
Check the light trap in the measuring cell for
residual water (circle).
Remove any residues with a cloth.