Instruction Manual OilGuard 2 W
Servicing
66
14809E/2
WORKSTEP
ADDITIONAL INFO / IMAGES
6.
Loosen the conduit gland (D) and remove the
outlet pipe (E) by pulling it downwards.
7.
Clean the removed parts or replace if neces-
sary.
When doing so, also inspect the seal in the
outlet cone (arrow).
8.
Insert the outlet pipe into the conduit gland
up to the stop and fasten loosely in place.
The outlet pipe should protrude approx.
20 mm into the measuring cell unit.
9.
Put the outlet cone (C) onto the outlet pipe
(E) up to the stop. When doing so, hold the
outlet pipe from below. Now press the outlet
cone together with the outlet pipe downward
up to the stop.