Instrument overview
Instruction Manual VisGuard 2
14162E/2
21
SICON (M):
Data
Values
Service voltage
24 VDC
Power consumption
5 W + photometer
Display
¼ VGA with touchscreen
Resolution: 320 x 240 pixels with 3.5" diagonal
Outputs/inputs
Outputs:
4 x 0/4 .. 20 mA, galvanically isolated up to max. 50 V relative to
ground and max. 500 Ω burden.
7 x digital outputs up to max. 30 VDC, freely configurable, 1 out-
put as de-energized closed relay.
Inputs:
5 x digital inputs up to max. 30 VDC, freely configurable
Interfaces
Ethernet, SD card (for logging, SW update, diagnostics),
Modbus TCP, optional Modbus RTU, Profibus DP, HART or
Profinet IO
Protection class
IP66
Weight
Approx. 0.6 kg
Dimensions
160 x 157 x 60 mm
Detailed dimension sheet according to Section 16
Housing material
ABS
Power Supply SP-C039
Data
Values
Service voltage
100 .. 240 VAC, 47 .. 63 Hz
Power consumption
Max. 25 W
Maximum operating
altitude
2,000 m above sea level
Protection class
IP66
Weight
1.1 kg
Dimensions
130 x 152 x 52 mm
Detailed dimension sheet according to Section 16
Housing material
PC