Mounting
Instruction Manual VisGuard 2
14162E/2
29
4.6
Installing the 230 VAC sample heater on the VisGuard 2
Proceed as follows to install the sample heater:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the splash guard from the sample inlet
of the VisGuard 2.
2.
Insert the sample heater up to the stop in the
sample inlet.
3.
Attach the splash guard to the sample heater.
4.
Establish the electrical connections to the
SIPORT 2 (arrow):
Terminal
1 (P)
2 (N)
3
Color
Brown
Blue
Yel-
low/green
Fuse protection is ensured via the line fu-
se of the SIPORT 2 power supply.