Operating the Calibrator / Micro Calibration Bath
© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
7.3.2 Manual control
Switching on the manual control:
It is possible to switch off the automatic control of the calibrator / micro calibration bath and to
achieve the desired temperature via manual control.
Press
the
P
key for approx 5 sec., the main menu opens. The last selected function ap-
pears on the upper display.
LED SET flashes on the lower display.
Press
the
or
key until
OPLO
appears.
OPLO
appears on the upper display.
LED SET flashes on the lower display.
Press
the
P
key to confirm
.
The current reference temperature appears on the
upper display.
The letter H and the currently set output capacity in
% appear on the lower display.
Press
the
key,
to
increase
the output capacity.
Press
the
key,
to
decrease
the output capacity.
PLEASE NOTE:
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys are
held pressed for at least one second, the value increases or decreases quickly and after
two seconds even more quickly; this means the desired value can be reached rapidly.
Switching off the manual control:
The manual control is switched on if the following display appears:
The current reference temperature appears on the
upper display.
The letter H and the currently set output capacity in
% appear on the lower display.
Switch the manual control off again by
pressing
the
P
key for approx. 5 sec., the main
menu opens.
OPLO
appears on the upper display.
LED SET flashes on the lower display.
Confirm switching on the automatic controller by
pressing the arrow key until the desired operating
mode is displayed and confirm this with the
P
key.
Menu manual control OPLO
Manual control OPLO setting display
Manual control OPLO setting display
OPLO display