S D B C - D K 3 U G
Rev. 0.3
35
7.2.9. Packet Error Test
Using the PER Lab Mode users may evaluate the following:
1. PER Sensitivity
2. FIFO mode using predefined packet structures (see datasheet for further details)
3. Receiver modem parameters:
i. Automatic
Frequency
Control
ii. Blocking
iii. Selectivity
4. Connect the receiver's input to the signal generator's output.
7.2.9.1. Test Method
1. Set frequency, modulation type, data rate and deviation parameters on the RF signal generator.
2. Program a predefined packet into the generator using the format:
(pr sync_word(2DD4H) + data + CRC).
3. Set the signal generator to external single trigger mode.
4. The software development board (SDB) generates a trigger on test point P1.4, this should be connected to the
RF signal generator's external trigger input. The P1.4 pin will enable the signal generator to send one packet for
each trigger.
Figure 49. Setup Screen (1 of 5)
1. Ensure Function is set to “Lab”
2. Set Lab Mode to “PER”
3. Press <GO> to move on from this screen.
Figure 50. Setup Screen (2 of 5)
1. Selections here should match those previously entered into the RF Signal Generator.
2. Press <GO> to move on from this screen.