VERTICALE
NAVIGATION
5
Assembly of Navigated Shaft Instruments
1)
Attach the corresponding Navigation tracker to the respective
Navigation Instrument (awl, probe, tap or SD Inner Shaft)* until
it snaps into place, make sure that the Navigation tracker is
securely connected to the respective instrument�
2)
Attach Silony handle according to your preference to the
proximal interface of the instrument shaft*�
3)
Additional steps for Navigated Screw Driver
The Navigated screw driver consists of a navigated driver shaft
in combination with modular components of the VERTICALE®
Screw driver VI-1130� Please refer to the instrument list
section at the end of this guide for combination options, e�g�
with SI-0024� For further (dis-)assembly instructions of the
modular screw driver VI-1130 or SI-0024, please refer to the
supplemental information provided with the instruments�
3.1)
Attach your desired outer sleeve option onto
the Basic Core�
3.2)
Instead of a non-navigated driver shaft, insert
the VI-7130 SD Inner Shaft into the respective
Screw driver Basic Core options�
* Further instrument options can be found in the instrument list
section at the end of this guide
** Further handle options can be found in the instrument list
section at the end of this guide
NOTE:
Do not use the same tracker type/geometry on
two different instruments without new registration
procedure� For registration instructions, please refer to
the registration section�
Fig� 1 Attaching Navigation tracker to navigated shaft
instrument�
GI-3111
Ratchet T-Handle**
Fig� 2 Assembling Navigated screwdriver shaft with Basic Core
of e�g� VI-1130�
NOTE:
If the tracker cannot be rigidly connected to the
respective Silony Navigation Instrument, utilize Silony
standard non-navigated surgical technique as described
in the VERTICALE® system guides�
Follow the instructions for NavLock™ tracker�
Assembly of instruments
Note:
For selection on tracker type or array size please
refer to the third-party’s user guides�