9.3. Air filter replacement / cleaning
1. Remove the filter's strips by removing the knurled screws.
2. Remove the filter cloth behind the grid by tilting the grid.
3. If necessary, clean the filter cloth using a vacuum cleaner and/or compressed air. The used
PU filters may be washed in water (allow them to dry properly).
4. Insert a clean filter of the correct length behind the grid.
5. Resnap the grid onto the blower's casing.
6. Remount the strip with the knurled screws and check that the entire air inlet is covered by
the filter cloth.
10. Faults
Problem Cause
Remedy
Red LED indicator lights up
(high voltage fault)
Short circuit in the anti-static
bar
Eliminate short circuit
Red and/or green LED
flashes
Short circuit in the anti-static
bar
Eliminate short circuit
Both LED indicators are off
No mains voltage present
Supply blower with mains
voltage
Poor ionisation
Fouled emitter points
Clean emitter points.
Insufficient air flow
Filter fouled / clogged
Clean or replace filter
Speed control set too low
Set the speed control to a
higher speed (Chapter 7)
Table 1 faults
If the solution does not solve the problem, please submit the Blow
ION
for repair. Please contact
SIMCO or a regional agent.
11. Repair
The Blow
ION
does not contain any parts which can be repaired or replaced by the user.
97.52.06.0910 Blow
ION
06/06
22